Calibration of a First-Generation Circular Variable Filter Spectrometer
Abstract
To permit accurate evaluation of previous airborne measurements, the AFCRL first-generation circular variable filter (CVF) spectrometer was calibrated in the AEROSPACE Research Chamber (7V). Testing consisted of calibration and linearity checks, angular response measurements, off-axis rejection measurements, and modulation transfer function determination. The calibration measurements defined the spectral radiance responsivity over a 4.0- to 13.0-micron wavelength range and the noise-equivalent radiance (NER) of the instrument.