APPARATUS FOR THERMOLUMINESCENCE. Part A
Abstract
A detailed description is given of an apparatus for thermoluminescence (above room temperature i.e. up to about 350 C), having the following peculiar features: the linear rate of warming is obtained through a cam-actuated Variac transformer; the light detection makes use of phase rectification; the apparatus is provided with a special furnace for recording glow current curves. The possibility of rendering flat the spectral response of the photomultiplier between 2000 and 6000 A through a composite filter is investigated and the method of calculating such a filter is developed. The performances of the apparatus were satisfactory.
Document Details
- Document Type
- Technical Report
- Publication Date
- Dec 01, 1961
- Accession Number
- AD0274007
Entities
People
- C. Cortese
- G. Bonfiglioli
- P. Brovetto