APPARATUS FOR THERMOLUMINESCENCE. Part A

Abstract

A detailed description is given of an apparatus for thermoluminescence (above room temperature i.e. up to about 350 C), having the following peculiar features: the linear rate of warming is obtained through a cam-actuated Variac transformer; the light detection makes use of phase rectification; the apparatus is provided with a special furnace for recording glow current curves. The possibility of rendering flat the spectral response of the photomultiplier between 2000 and 6000 A through a composite filter is investigated and the method of calculating such a filter is developed. The performances of the apparatus were satisfactory.